Capacitance
Range
1)
500.0nF, 5.000μF, 50.00μF,
500.0μF, 3000μF
1) Additional 50.00nF range accuracy is not specified
2) Accuracies with film capacitor or better
3) Specified with battery voltage above 2.8V (approximately half full battery).
Accuracy decreases gradually to 12% rdg at low battery warning voltage of
approximately 2.4V
Transient protection:
DC μA
Range
Accuracy
400.0μA
±( 2.0% rdg + 4 digits)
2000μA
±( 1.2% rdg + 3 digits)
Transient protection:
AC μA (50Hz ~ 500Hz)
Range
Accuracy
400.0μA
±( 2.0% rdg + 5 digits)
2000μA
±( 1.2% rdg + 5 digits)
Transient protection:
Type-K Temperature
Range
-20 °C ~ 300 °C
301 °C ~ 537 °C
-4 °F ~ 572 °F
573 °F ~ 999 °F
Type-K thermocouple range & accuracy not included
Accuracy
±( 3.5% rdg + 6 digits)
6.5 kV (1.2/50 μs surge)
6.5 kV (1.2/50 μs surge)
6.5 kV (1.2/50 μs surge)
Accuracy
±( 2% rdg + 3 °C)
±( 3% rdg + 3 °C)
±( 2% rdg + 6 °F)
±( 3% rdg + 6 °F)
2) 3)
Burden Voltage
2.8mV/μA
2.8mV/μA
Burden Voltage
2.8mV/μA
2.8mV/μA
13