Application Programs Counting; Weighing In Percent - Boeco BBL Serie Manual De Instrucciones

Balanzas electronicas de precisión para metales preciosos
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Application Programs
Counting
Purpose
With the Counting program you can determine the number
of parts that each have approximately equal weight.
Available Features
– Store the current weight value to have it loaded as the
preset reference sample quantity next time you initialize
the Counting application
– The reference sample quantity can be changed in the
operating menu: See "Configuration"
– The average piece weight is automatically output via the
optional data interface port after initialization, if the menu
code for "Printout with data ID codes" is set
– Press [FUNCTION] to toggle the display between piece
count and weight
Factory Settings
Reference sample quantity: 10 (menu code
Preparation
§ Configure the Counting application in the operating menu:
See "Configuration"
2. 1. 4
Menu code
§ Reference sample quantity:
3. 3. 1
Menu code
3. 3. 2
Menu code
3. 3. 3
Menu code
3. 3. 4
Menu code
3. 3. 5
Menu code
See also "Configuration"
6
All manuals and user guides at all-guides.com
3. 3. 2
)
Counting
5 pcs
10 pcs
20 pcs
50 pcs
100 pcs
Example
Determine an unknown piece count; weigh the preset
reference sample quantity
Settings (changes in the factory settings required for
this example):
Menu: Application program: Counting (
Reference sample quantity: 20 pcs (
Step
Key (or instruction) Display
1. Turn on the balance
[ON/OFF]
t
2. Tare the balance
3. Display the reference
[FUNCTION]
sample quantity
> 2 sec.
(here: 20 pcs)
4. Place the reference sample
quantity (20 pcs) on the
balance (here: 66 g)
5. Start the application;
[FUNCTION]
if the print format is set
to include data ID codes,
the following is printed
6. Weigh uncounted parts
(here: 174 pcs)
7. Display weight
[FUNCTION]
8. Display quantity
[FUNCTION]
9. Unload the balance
10. Delete the reference
value
[CAL/CF] > 2 sec.
11. Repeat the procedure
starting from Step 6,
if desired.

Weighing in Percent

Purpose
This application program allows you to obtain weight read-
outs in percent which are in proportion to a reference weight.
Available Features
– Store the current weight value to have it loaded as
the preset reference percentage next time you initialize
the Weighing in Percent application
– The reference percentage can be changed in the
operating menu: See "Configuration"
– The reference percentage is automatically output via the
optional data interface port after initialization, if the menu
code for "Printout with data ID codes" is set
– Press [FUNCTION] to toggle the display between
percentage and weight
Factory Settings
Reference percentage: 10 (menu code
2 1 4
)
3. 3. 3
)
0.0
g
ref 20
(briefly)
66.0
+
g
20
+
pcs
wRef
+
3.300 g
174
+
pcs
574.2
+
pcs
174
+
pcs
0
pcs
3. 3. 2
)

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido