Application Programs
Counting
Purpose
With the Counting program you can determine the number
of parts that each have approximately equal weight.
Available Features
– Store the current weight value to have it loaded as the
preset reference sample quantity next time you initialize
the Counting application
– The reference sample quantity can be changed in the
operating menu: See "Configuration"
– The average piece weight is automatically output via the
optional data interface port after initialization, if the menu
code for "Printout with data ID codes" is set
– Press [FUNCTION] to toggle the display between piece
count and weight
Factory Settings
Reference sample quantity: 10 (menu code
Preparation
§ Configure the Counting application in the operating menu:
See "Configuration"
2. 1. 4
Menu code
§ Reference sample quantity:
3. 3. 1
Menu code
3. 3. 2
Menu code
3. 3. 3
Menu code
3. 3. 4
Menu code
3. 3. 5
Menu code
See also "Configuration"
6
All manuals and user guides at all-guides.com
3. 3. 2
)
Counting
5 pcs
10 pcs
20 pcs
50 pcs
100 pcs
Example
Determine an unknown piece count; weigh the preset
reference sample quantity
Settings (changes in the factory settings required for
this example):
Menu: Application program: Counting (
Reference sample quantity: 20 pcs (
Step
Key (or instruction) Display
1. Turn on the balance
[ON/OFF]
t
2. Tare the balance
3. Display the reference
[FUNCTION]
sample quantity
> 2 sec.
(here: 20 pcs)
4. Place the reference sample
quantity (20 pcs) on the
balance (here: 66 g)
5. Start the application;
[FUNCTION]
if the print format is set
to include data ID codes,
the following is printed
6. Weigh uncounted parts
(here: 174 pcs)
7. Display weight
[FUNCTION]
8. Display quantity
[FUNCTION]
9. Unload the balance
10. Delete the reference
value
[CAL/CF] > 2 sec.
11. Repeat the procedure
starting from Step 6,
if desired.
Weighing in Percent
Purpose
This application program allows you to obtain weight read-
outs in percent which are in proportion to a reference weight.
Available Features
– Store the current weight value to have it loaded as
the preset reference percentage next time you initialize
the Weighing in Percent application
– The reference percentage can be changed in the
operating menu: See "Configuration"
– The reference percentage is automatically output via the
optional data interface port after initialization, if the menu
code for "Printout with data ID codes" is set
– Press [FUNCTION] to toggle the display between
percentage and weight
Factory Settings
Reference percentage: 10 (menu code
2 1 4
)
3. 3. 3
)
0.0
g
ref 20
(briefly)
66.0
+
g
20
+
pcs
wRef
+
3.300 g
174
+
pcs
574.2
+
pcs
174
+
pcs
0
pcs
3. 3. 2
)