G5000
Safety Manual
In hazardous area only use "Temperature Probe GF5.2" in Connector B.
Connector C (Uo=10V,lo=5mA,Po=50mW,Ci=0,Li=0,Co=100uF,Lo=1000mH),
Connector B (Uo=5V,lo=6mA,Po=7mW,Ci=0,Li=0,Co=100uF,Lo=1000mH)
•
The safe area apparatus that is to be connected to the USB Port shall be a Safety Extra Low
Voltage (SELV) or Protective Extra Low Voltage (PELV) circuit.
•
Only a QED Environmental Systems Ltd battery pack part number 20087 or 2011113 is
permitted as a replacement. This battery pack shall only be changed in a safe area.
•
Only Battery Charger type GF3.9 shall be used to recharge the batteries via Connector 'B'.
•
If the equipment is likely to come into contact with aggressive substances, e.g. acidic liquids or
gases that may attack metals, or solvents that may affect polymeric materials, then it is the
responsibility of the user to take suitable precautions, e.g. regular checks as part of routine
inspections or establishing from the material's data sheet that it is resistant to specific
chemicals that prevent it from being adversely affected, thus ensuring that the type of
protection is not compromised.
•
The relative pressure range is +/-500 mbar. Note, however, that the input pressure for the
GA5000 should not exceed +/- 500 mbar relative to atmospheric pressure, and the input
pressure for the GEM5000 and the BIOGAS 5000 should not exceed +/- 250 mbar relative to
atmospheric pressure, output pressure should not exceed +/- 100 mbar relative to
atmospheric pressure.
Instructions for safe use of Battery charger
•
Please read instructions before using the charger.
•
For indoor use only.
•
Do not use to for charging non-rechargeable batteries.
•
The battery pack has six nickel-metal hydride cells, normal taking 6 hours to charging time.
•
When the battery is worn out, dispose of it under the ordinance of each local government.
•
The applicance is not intented fro use by persons or children with reduced physical, sensory or
mental capability, or lack of experience, unless supervised.
Bluetooth
This device contains: FCC ID: PVH092102 IC: 5325A-092102
www.qedenv.com
MAXIMUM NON-HAZARDOUS SUPPLIES:
Connector A - Um=6V Connector B - Um=10.1V
Page 9 of 44
OM5KSAFETY Issue 11
© QED Environmental Systems Ltd.