Vacuum system SH820G / SR820G SH840G / SR840G
Translation of Originial Operating Instruction, KNF 325562-325564 06/21
NOTICE
Status display (see 10/Fig. 2 and Fig. 3)
à
Illuminates green if the vacuum system was switched on
via the power switch.
à
Illuminates blue if the vacuum system was also switched
on via the rotary/push knob.
Vacuum system running:
The higher the speed is set, the brighter the status display
lights up.
à
Illuminates red if there is a fault:
Signal duration
100% ON (continuous light)
50% ON; 50% OFF
90% ON, 10% OFF
Tab.14: Fault signal via status display
For further information, see Chapter 11 Troubleshooting
[} 60].
Damage may occur due to incorrect or
improper transport
If the collection flasks remain mounted
during transport of the vacuum system,
they can be damaged by careless han-
dling when the vacuum system is set
down.
à
Remove the collection flasks be-
fore transporting the vacuum sys-
tem.
à
If there are liquids in the collection
flasks, empty them or dispose of
the liquids in an environmentally
friendly manner.
à
Store the collection flasks in a safe
location.
à
Reassemble the collection flasks
after transport.
Fault type
Drive blocked
Temperature too high (drive)
Other fault