Device stabilization in process.
Solution: Wait until stops flashing.
Battery is low and no more readings are possible.
Solution: Replace the battery.
Reading before device stabilization.
Solution: Wait until stops flashing.
The ambient temperature is not within the range between 50°F ~104°F (10°C and 40°C).
Solution: Allow the thermometer to rest in a room for at least 30 minutes at room temperature:
50°F ~104°F (10°C and 40°C).
Error 5~9, the system is not functioning properly.
Solution: Unload the battery, wait for 1 minute and then turn on.
Temperature taken is higher than 108.0°F (42.2°C).
Solution: Check the integrity of the hygiene cover and take a new temperature reading.
Temperature taken is lower than 93.2°F (34°C).
Solution: Make sure the hygiene cover is clean and take a new temperature reading.
Device can not be powered on to the ready stage.
Solution: Insert a new battery.
Clinical repeatability 0.18°C newborn to 1 year old, 0.17°C 1-5years old, 0.15°C 5years+
This product comes with a 12-month warranty.
Manufacture Date as the serial number (please open the battery cover, it is shown on the inside of
the device.) Ex.SN:E912A000001, the first "E" is External, the second number "9" is the
manufacture year 2009, the third and the fourth number "12" is the manufacture month, the
remaining figures are the serial number. Our batch code is XXXXVT.
BF type applied part.
TROUBLESHOOTING