RESULT
CAL
Mode
Zero
Test
CAL
RESULT + x
On
Off
:
:
Store
Date
Cal
RESULT
Time
Cal
Zero
Mode
Test
On
Off
SEL
CAL
SEL
cAL
Mode
SEL
CAL
On
Off
14
The result is shown in the display, alternating with CAL.
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
By pressing the [ON/OFF] key, the new correction factor is calculated
and stored in the user calibration software.
Confirmation of calibration (3 seconds).
Factory calibration reset
A user calibrated method is indicated by an arrow while the test result
is displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST] key
and hold.
Switch the unit on using the [ON/OFF] key.
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
The following messages will appear in turn on the display:
The factory setting is active.
(SEL stands for Select)
or:
Calibration has been set by the user.
(If the user calibration is to be retained, switch the unit off using the
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:
Switch the unit off using the [ON/OFF] key.
VWR ver 1 23.01.2018