VWR MCP 100 Manual De Instrucciones página 23

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 59
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Ozone may show lower results. To avoid any measure-
ment errors, only use glassware free of Chlorine demand.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample:
When preparing the sample, the lost of Ozone, e.g. by pipetting or shaking, must be
avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents
therefore contain a buffer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before
the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range:
Concentrations above 2 mg/l Ozone can lead to results showing 0 mg/l. In this case,
the water sample must be diluted with water free of Chlorine and the measurement
repeated.
5. Oxidising agents such as Bromine, Chlorine etc. interfere as they react in the same way
as Ozone.
30144.294
31627.290
85413.601
84208.600
VWR ver 1 23.01.2018
SULPHURIC ACID 0.5 MOL/L 1N AVS TITRINORM
SODIUM HYDROXIDE 1 MOL/L AVS R.PE/USP/NF
PH TEST STRIPS 4.5-10.0 DOSATEST
REFERENCE STANDARD KIT RSK CHLORINE MCP 100
23

Publicidad

Tabla de contenido
loading

Tabla de contenido