Troubleshooting
EN-
Error
US
X-ray image has strips on
image
Bright stripes in the scanning
window
Horizontal, grey lines in the X-
ray image, extending beyond
the left and right image edge
X-ray image is stretched
lengthwise with bright, hori-
zontal stripes
52
Possible cause
PSP has been pre-exposed, e.g.
by natural radiation or stray X-
ray radiation
Parts of PSP exposed to light
during handling
PSP dirty or scratched
Too much incident ambient light
during the scanning process
Transport slipping
Wrong barrier envelope or PSP
used
Remedy
❯ If the PSP has been stored for
more than one week, erase
the PSP prior to use.
❯ Do not expose exposed PSPs
to bright light.
❯ Scan image data within half
an hour after the exposure.
❯ Clean the PSP.
❯ Replace scratched PSP.
❯ Darken the room.
❯ Turn the unit such that no light
is directly incident on the input
unit.
❯ Clean the transport mecha-
nism, replace the transport
belts if necessary.
❯ Only use original accessories.
2151100018L40 2012V016