MANUAL DE INSTRUCCIONES
Electric risk
Risk of high
temperature contact
Reference code:
Dimensions: height/width/depth (mm)
Supply voltage:
Electrical power (W):
Weight: (Kg)
External fuse:
Temperature range: (No refrigeration)
Display resolution:
Description
It does not heat
Temperature reading does
not match the external ther-
mometer
J
.P. SELECTA s.a.u.
CODIGO 80300 REV C
7. MAINTENANCE & CLEANING
MAINTENANCE:
Before operating inside, disconnect from the mains.
The manipulation of the equipment internal electronic circuits by unauthorized
personnel can cause irreparable damage. Take it to one of ICC's authorized
technical services.
CLEANING:
To clean the accessible parts of the equipment, use a cleaner or a similar product.
Do not use aggressive solvents. To remove lime, you can use vinegar a bit lowered.
8. TECHNICAL SPECIFICATIONS
230 VAC 50 - 60 Hz
9. ERROR MESSAGES & TROUBLESHOOTING
P.FAIL
Err
will appear in green on the top line of the display.
(Note: if you turn off the device without pressing STOP, this error will appear when
the device is restarted)
TSEG
The safety thermostat is open.
SENS
Open temperature probe.
The control does not detect the temperature probe.
HEAT
The resistance does not work.
The equipment detects that the temperature does not increase.
Cause
The safety thermostat has skipped
Heating element is melted
Temperature imbalance
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
ABRIL 2019
(Sujetas a modificaciones sin previo aviso)
80013
280/180/190
2060
4
Replaceable (Type: 10x5mm)
Ambient Tª + 5ºC ...
0.1ºC
Power failure. The equipment continues with the process.
Resolution
Press the thermostat reset. See 9.1
Contact the technical service to replace
See 9.2 on how to act for adjustment
80015
115 VAC 50 - 60Hz
1060
150ºC
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 9