The values of tube voltage and current are configured automatically according to the
patient's gender / age group and X-ray intensity. Click the UP/DOWN arrow to adjust
kVp and mA. The dose is adjustable by ±1 kVp and ±1 mA respectively.
9.
Click CONFIRM button when exposure parameter setting is completed.
10.
Guide the patient to the equipment.
PaX-i Plus User Manual
When you click CONFIRM button,
The Rotating Unit will move to its scanning position.
READY button will be activated. (This means that the
equipment is ready for X-ray exposure.)
Three Laser Beams (Vertical Beam, Horizontal Beam and
Canine Teath Beam) will be activated.
- The Laser Beams are turned off automatically after 20
minutes or when READY button is clicked.
The DAP (Dose Area Product), Scan Time and Exposure
Time will be displayed below Patient Information window.
3. Acquiring PANO Images