Switch on the BasicMes-2 and connect to PC
•
Change to menu item "Calibration" in main menu of
the BasicMes-2
•
Set access code to "3318"
•
On PC select menu "Device" and menu item
"Firmware update"
•
Start process on PC by click onto "Update"
•
Choose firmware file
•
Confirm update on the BasicMes-2
•
Update starts automatically and takes about 3
minutes. The display of the BasicMes-2 remains blank
during this time
•
The BasicMes-2 restarts after successful update.
USB connection can now be removed
7
EU Declaration of conformity
The product: Resideo VM242A BasicMes-2
handheld flow measuring computer
conforms with the essential protection requirements which
are set out in the directives of the European Council for to
adapt the legal provisions of the Member States in respect of
electromagnetic compatibility (2004/108/EG).
The following standards were availabled of for to evaluate the
product in respect of electromagnetic compatibility:
EN61326-1:1997 + A1:1998+A2:2001
This declaration is based on the measurements of a third
party.