VIII. PRECISION
ACCORDING TO
ISO 80601-2-56
The THERMOFLASH® LX-26E can take temperature measurements below 30°C /
89°F or above 42.9°C / 109.22°F. Outside these temperature ranges, precision is
not guaranteed.
Longevity of use
The ThermoFlash® Premium is guaranteed for about 40,000 readings.
IX. ACCESSORIES SUPPLIED
User manual, LR6 (AA)*2PCS.
X. TROUBLESHOOTING
If you experience any of these problems using your THERMOFLASH® Premium,
please refer to this troubleshooting guide to resolve the problem. If the problem
persists, please contact our customer service at +33 3 88 78 85 24.
- The display indicates temperatures above 95°: the temperature is indicated in
Fahrenheit. Change the unit of measure to Celsius.
- The screen displays temperatures below 30°C / 86°F : to measure body
temperature, the mode BODY must be selected. If the thermometer is in SURFACE
or ROOM mode, the displayed reading indicates the external temperature
released by your body.
- The screen displays the message HI : the analysed temperature is above the
range of the selected mode, i.e. greater than 42.9°C / 109.2°F in BODY mode, or
90°C / 194°F in SURFACE mode.
- The screen displays the message LO : The analysed temperature
is below the measuring range of the selected mode, i.e. less than
30°C / 86°F) in BODY mode and less than 0°C / 32°F in SURFACE mode.
This message displays in various cases - below are a list of the main cases:
30°C - 34,9°C =
± 0,3°C
35°C - 42°C = ± 0,2°C
42,1°C - 43°C = ±0,3°C
86°F - 94,82°F =
±0,5°F
95°F–107.6°F: ±0.4°F
107,78°F - 109,4°F =
±0,5°F
21