Recommended Separation Distance - Biosynex ThermoFlash Premium Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
810
870
930
1720
1845
1970
2450
5240
5500
5785
Electromagnetic Environment - Guidance
RF wireless communications equipment should be used no
closer to any part of the device, including cables, than the rec-
ommended separation distance calculated from the equation
applicable to the frequency of the transmitter.

Recommended separation distance

Where P is the maximum output power rating of the ransmit-
ter in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in meters (m). Field
strengths from fixed RF transmitter, as determined by an electro-
magnetic site survey, should be less than the compliance level in
each frequency range. Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1: These guidelines may not apply in all situations. Electro-
magnetic propagation is affected by absorption and reflection from
structures, objects and people.
WARNING
• This device should not be used in the vicinity or on the top of other
electronic equipment such as cell phone, transceiver or radio con-
trol products. If you have to do so, the device should be observed to
verify normal operation.
• The use of accessories and power cord other than those specified,
with the exception of cables sold by the manufacturer of the equip-
ment or system as replacement parts for internal components, may
result in increased emissions or decreased immunity of the equip-
ment or system.
2
0.3
2
0.3
2
0.3
0.2
0.3
28
28
28
28
28
28
9
9
E= [ 6
] √
P
d
77