EMC TABLES
Recommended separation distances between portable and mobile RF
communications equipment and the TIE-210 Thermometer
The TIE-210 Thermometer is intended fo use in an electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of the TIE-210 Thermometer can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
TIE-210 Thermometer as recommended below, according to the maximum
output power of the communications equipment.
Rated
maximum
output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance according to frequency of
150 kHz to
80 MHZ
d=1.2 √P
0.12
0.38
1.2
3.8
12
21
transmitter m
80 MHz to
800 MHZ
d=1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz to
2.5 GHz
d=2.3 √P
0.23
0.73
2.3
7.3
23