7 Maintenance

Disconnect power cord before cleaning the light booth or replacing lamps.
Lamps are hot after use! Let them cool down first.
Cleaning
To clean the neutral gray surface or diffuser, use a soft, damp cloth with a wa-
ter based mild soap or detergent. Allow to air dry. Release the latch to reach
the inner side of the diffuser.
Lamp replacement procedure
Remove the front panel of the light booth. Release diffuser to reach lamps. To
protect the lamps, only touch them with gloves.
Lamp type
Daylight (D65)
Horizon (HZ) /
Incandescent (A)
CWF / TL84 / U30 /
Ultra-violet (UV)
301 300 039 - 2109
Lifetime
Replacement note
600 hrs
Please contact our technical service for
replacement.
600 hrs
Please push one end of the lamp into
the spring-loaded socket and release the
other end.
After replacement perform Self-Calib-
ration as described on the next page.
2000 hrs
Please rotate the tube 90 degrees to
loosen it.