Self-Calibration of A and HZ lamps
After replacing A or Horizon lamps, you MUST use the Self-Calibration func-
tion to calibrate them.
1.
Go to SETUP > ADMIN > RLAMP.
2.
Use ▲ and ▼ to highlight either A or HZ.
3.
Press ENTER to access lamp replacement screen.
4.
Input lamp parameters "Remaining Hours, Remaining Cycles and Serial
number".
5.
Use ▲ and ▼ to highlight "Self-Calibration" and press ENTER.
The self-calibration process takes 5 minutes to finish. The button on the screen
switches from "Calibrating..." to "Complete".
Press ENTER to return to the lamp replacement screen.
12