is displayed soon after
pressurization is complete, and
measurement begins automatically.
* If the unit judges that pressurization
is insufficient, it automatically
repressurizes.
blinks each time a pulse is
detected, and the pressure value
is displayed on the diastolic blood
pressure display.
Once measurement is complete, the
cuff deflates and the measurement
results (systolic/diastolic blood
pressure values, pulse) are displayed.
* Systolic/diastolic blood pressure
values (two seconds) and pulse (one
second) are displayed alternately.
* Provided there is no error in the measurement results, they are stored
automatically.
The Body Movement Indicator is displayed on the measurement
results display if you move your hand or arm during measurement,
causing a substantial pressure change to be detected.
• If the Body Movement Indicator is displayed, measure your blood
pressure again.
If an irregular heartbeat is detected during measurement, the
Irregular Heartbeat (IHB) Indicator is displayed after measurement
is completed.
An irregular heart beat is defined as a heart beat that varies by
±25% from the middle of 5 heart beat intervals during the blood
pressure measurement.
• Measurement may not be correct if your heartbeat fluctuates greatly
during measurement. If the Irregular Heartbeat (IHB) Indicator is
displayed, measure your blood pressure again.
Caution
MEASURING YOUR BLOOD PRESSURE
If the Irregular Heartbeat (IHB) Indicator appears
frequently, you are advised to consult a doctor about
your health condition.