Recommended separation distance between portable and mobile radio-communication devices and the monitor
The AS-36 BR surgical aspirator is intended to operate in an electro-magnetic environment where RF irradiated interferences
are under control. The client or operator of the AS-36 BR device can help prevent electro-magnetic interference by keeping a
minimum distance between the portable and mobile RF communication devices (transmitters) and the AS-36 BR device, as
recommended below, in relation to the radio-communication maximum output power.
Maximum nominal
output power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people.
This product is guaranteed for a period of 24 months from the date of purchase. The warranty includes the repair or replacement
of defect spare parts free of charge, if the defect has been clearly described by the customer and determined by technical service.
Inspections on the part of the seller, performed at the request of the customer and intended to determined wjether the device is
fully functional, are not covered by the free-of-charge warranty service. This service will be charged to the customer depending on
the effort required. The consumables components are not subject to warranty. Consumable components are silicon tubes, filters,
seals, conical adaptor and suction catheter. Also exluded from warranty is all damage resulting from improper handling, wilful
damage or improper care of the device.
The warranty shall expire if repairs and servicing are not carried out by technical service.
UNDER NEW EUROPEAN RULES, CA-MI REQUIRES THE FOLLOWING PROCEDURES TO BE CARRIED OUT TO PROTECT THE
INSTRUMENT AND THE SAFETY OF ALL WHO COME IN CONTACT WITH IT.
Before returning an instrument for repair, the external surfaces and all accessories MUST be carefully disinfected with a cloth
soaked in methylated spirits or hypochlorite-based solution. The instrument and accessories should then be placed in a bag with a
note outlining the disinfection undertaken.
Failure to follow this procedure will result in the instrument being returned to the purchaser unrepaired.
Instruments returned for repair MUST be accompanied by a description of the problem. CA-MI will not be responsible for damage
caused through improper use. To avoid such damage, please read the instruction carefully.
Where CA-MI determines that an instrument is faulty, a replacement will be provided only if a SALES RECEIPT and STAMPED
GUARANTEE are provided. CA-MI will not be responsible for damage accessories. These may be replaced at the customer's
expence.
Separation distance from the frequency transmitter (m)
150 kHz to 80 MHz
d = [3.5 / V
]
P
1
0,12
0,38
1,2
3,8
12
WARRANTY CONDITIONS
RULES FOR RETURNING AND RAPAIRING
80 MHz to 800 MHz
d = [12/E
]
P
1
0,12
0,38
1,2
3,8
12
22
GB
800 MHz to 2.5 GHz
d = [23/E
]
P
1
0,23
0,73
2,3
7,3
23