6
Commissioning the unit
6.1 Before starting work
Before starting surgery, the HBS x-touch 300 argon BiSeal
supply cable to the mains socket with PE conductor. The current cable may only be attached or
removed when the device is switched off. The connecting socket for the power cable is located at the
back of the housing.
Foot switches are connected at "1mono/bi" or "2 mono/bi + auto". The terminals are located at the
back of the housing.
Handles and electrodes are connected correspondingly at the front of the housing to
"MONOPOLAR1", "MONOPOLAR2" or "BIPOLAR".
6.2 Connection of argon
!
The HBS x-touch 300 argon BiSeal
transmission system.
Gas flow from gas transmission system
•
Close the hose at one of the two connection points (D or G).
Use of gas cylinders
•
Secure the gas cylinder (A) with the provided fixture (B)
•
Do NOT remove the identification label
•
Mount an original HEBU pressure relief valve HEBU (C) at the
gas cylinder (article number: HF 9530-02)
®
Please observe the following warning remarks with regard to
argon-supported coagulation and cutting
Before first using the electrosurgical unit, we recommend
familiarizing employees with the operating characteristics of the
different operating modes and power settings by carrying out
practice exercises using fresh meat (beef, pork).
Use ultra-pure argon gas with a purity level of 99.996 % or
higher. Using gas with a lower purity can place the safety of
®
is able to work both with gas from a cylinder or from a gas
®
electrosurgery unit is connected via the
patients in jeopardy.