7. Warranty / Procedure
in Case of Repair
The warranty period for Bluephase PowerCure is 3 years
from the date of purchase (battery: 1 year). Malfunctions
resulting from faulty material or manufacturing errors
are repaired free of charge during the warranty period.
The warranty does not provide the right to recover any
material or non-material damage other than the ones
mentioned. The apparatus must only be used for the
intended purposes. Any other uses are contraindicated.
The manufacturer does not accept any liability resulting
from misuse and warranty claims cannot be accepted in
such cases. This is particularly valid for:
• Damage resulting from improper handling, especially
incorrectly stored batteries (see Product Specifi cations:
Transportation and storage conditions).
• Damage to components resulting from wear under
standard operating conditions (e.g. battery).
• Damage resulting from external influences, e.g. blows,
drop to the floor.
• Damage resulting from incorrect set-up or installation.
• Damage resulting from connecting the unit to a power
supply, the voltage and frequency of which do not
comply with the ones stated on the rating plate.
• Damage resulting from improper repairs or
modifications that have not been carried out by
certified Service Centres.
In case of a claim under warranty, the complete
apparatus (handpiece, charging base, battery, power
cord and power pack) must be returned, carriage paid, to
the dealer or directly to Ivoclar Vivadent, together with
the purchase document. Use the original packaging with
the corresponding cardboard inserts for transportation.
Repair work may only be carried out by a certified Ivoclar
Vivadent Service Centre. In case of a defect that cannot
be rectified, please contact your dealer or your local
Service Centre (see addresses on the reverse side). A
clear description of the defect or the conditions under
which the defect occurred will facilitate locating the
problem. Please enclose this description when returning
the apparatus.

8. Product Specifications

Ivoclar Vivadent Polywave
Light source
Wavelength range
385–515 nm
Light intensity
3sCure program: 3,050 mW/cm
Turbo program: 2,100 mW/cm
High Power program: 1,200 mW/cm
PreCure program: 950 mW/cm
Operation
3 min on / 7 min off (intermittent)
Light guide
10>9 mm, autoclavable
Signal transmitter
acoustic every 10 seconds and every time the
start/stop button or the time / program
selection button is activated or the anti-glare
protection is enabled or the curing process is
aborted and 30-second waiting time after the
3sCure program has been used twice in
succession
Dimensions of the
L = 170 mm, B = 30 mm, H = 30 mm
handpiece
Weight of the handpiece
135 g (including battery and light guide)
Operating voltage
3.7 VDC with battery
handpiece
5 VDC with power pack
Operating voltage of the
5 VDC
charging base
Power supply
Input: 100–240 VAC, 50–60 Hz max 1 A
Output: 5 VDC / 3 A
Manufacturer: EDAC POWER ELEC.
Type: EM1024B2
Operating conditions
Temperature +10 °C to +35 °C
Relative humidity 30 % to 75 %
Ambient pressure 700 hPa to 1060 hPa
Dimensions of the
D = 110 mm, H = 55 mm
charging base
Weight of the charging
145 g
base
Charging time
Approx. 2 hours (with the battery empty)
Power supply of the
Li-Ionen battery
handpiece
Transportation and
Temperature –20 °C to +60 °C
storage conditions
Relative humidity 10 % to 75 %
Ambient pressure 500 hPa to 1060 hPa
The curing light has to be stored in closed,
roofed rooms and must not be exposed to
severe jarring.
Battery:
– Do not store at temperatures above 40 °C /
104 °F (or 60 °C / 140 °F for a short period).
Recommended storage temperature
15–30 °C / 59–86 °F
– Keep the battery charged and store no longer
than 6 months.
Delivery form
1 Charging base with power cord and power
pack
1 Handpiece
1 Handpiece support
1 Light guide 10>9 mm
1 Anti-glare shield
3 Anti-glare cones
1 Pack of sleeves (1x 50 pcs)
1 Instructions for Use
LED
®
2
± 10 %
2
± 10 %
± 10 %
2
2
± 10 %
19

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido