be in the center of the clamp aperture because the measurement may be affec-
ted by the position within the clamp aperture of the conductor being measured.
4. Press the opening lever until the UNLOCK indication disappears, and check that
the sensor head is properly closed.
If the sensor head is not properly closed, accurate measurement is not possible.
The current consumption of clamp-on probes depends on the current to be
measured. Confirm that the total current consumption of the probes does not
exceed the rated output current of the R&S RT-ZA13 probe power supply.
Fig. 3-10: Current consumption* vs. current to be measured (typical)
*The sum total of a positive and negative current consumption
User Manual 1409.7814.02 ─ 01
Putting into Operation
Connecting the Probe to the DUT
23