Descargar Imprimir esta página

Interconnections; Earth (Ground) Connection - Agilent Technologies TwisTorr 804 FS Serie Manual De Istrucciones

Publicidad

17
Technical Information

Interconnections

Interconnections

The following figure shows the TwisTorr 804 FS interconnections.
Figure 36

Earth (Ground) Connection

It is required to connect the pump to protective earth with a M5
screw (marked thread in the pump envelope) using a green/yellow
wire of size AWG 16. Please ensure a reliable connection using an M5
screw and a snap ring or washer disc of the same metal as the pump
body to prevent corrosion. The screw must be fixed with a 2 Nm
torque.
The impedance between the pump-body and the earth connection
point must be < 0.1 Ω.
306/344
TwisTorr 804 FS User Manual / 87-901-052-01

Publicidad

loading