1.9.Manteinance
- You should clean the still regularly. So, in
this way, the results will be better and you
can get a still more durable.
- The cleaning periodicity depends on the
salts which water supply contains.
- You should clean the filters of the tubes
whenever you clean the still.
Steps to follow:
1. Put acetic acid, hydrochloric acid or
whichever another product into the boiler
through cleaning mouth (piece 5). The water
should be hot.
2. You have to wait for the attack to the
residues and empty the boiler through the
same cleaning mouth (piece 5).
3. Rinse with clean water once or twice.
4. Fill with water until you see the water
level is on the heater.
Rev. 020913
1.10.
Spare parts
Description
Complete water still
1
Nut, nº 42
4
Condenser
5
Cap GL-32
6
Level
7
Nut GL-25, hole 15 mm
8
Flange,
9
Boiler
12
Support with automatic device and switch
14
Heater
702 724
Pag: 10
Reference
720 720
721 642
919 916
162 045
919 926
161 836
702 724
919 913
702 004
721 548