Technical Information
Optional Serial Port

RS 232 Communication Port

Communication serial port connections and minimum connection
configuration are shown in the following figures. The communication
port mating connector is supplied with the RS 232 PCB
(AMP/Cannon or equivalent 9-pin "D" type male connector). The
external cable (not supplied) between the host computer and the
controller requires some crossed wires so that signals are connected
correctly.
For example, the Transmit data signal from controller (pin 2) must be
connected to the host computer's Receive data line (pin 3) and vice
versa. Consult the host computer's instruction manual for its serial
port connections.
Agilent cannot guarantee compliance with FCC regulations for radiated emissions
NOTE
unless all external wiring is shielded, with the shield being terminated to the
metal shroud on the O-subconnector. The cable should be secured to the
connector with screws.