issue
Air from flow
generator seems to
be too warm.
Nasal, sinus or ear
pain, runny nose
Heater setting is
not visible on
display
Bilevel changes
pressure too early or
too late (not
following breathing)
SpECiFiCaTiONS
Cpap
Size ........................................................................................................... 4.2"H x 6.5"W x 6.9"D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ...................................................................................................................... 100-240V~, 50/60 Hz
Electrical Requirements DC .................................................................................................................. 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Bilevel Pressure Range ...........................................................................................................................................3-25 cmH
Operating Temperature Range ....................................................................................................... 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range .......................................................................................................0% to 95% RH non-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 8,500 feet (2600 m)
Storage & Transportation Temperature Range ..................................................................... -40°F to +158°F (-40°C to 70°C)
Storage & Transportation Humidity Range ...............................................................................0% to 95% RH non-condensing
Maximum Limited Pressure ...........................................................................................................30 cmH
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
SmartCode
Technology is protected under US patent 8649510
®
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10 mgH
Size: ......................................................................................................2.6" H x 6.3" W x 8.4" D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................305°F (152°C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: ...............................................................................84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Temperature Range: ....................................................................................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range: .......................................................................................................0 to 95% R.H. non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA
Filter Specifications
Standard Filter .........................................................................................................................................> 3.0 micron particles
Optional Fine Particle Filter .....................................................................................................................> 0.3 micron particles
SE-DV56-1
possible Cause
1. Air filters are dirty.
2. Air inlet port is blocked.
3. Room temperature is too high.
4. Device is located near a heat source.
5. Humidifier (optional) heater setting is too
high.
You may have a reaction to air flow pressure. Discontinue use and contact physician.
1. The flow generator is not fully seated onto
the humidifier cradle.
2. Your CPAP device is not equipped with the
optional humidifier.
1. Trigger Sensitivity is not set correctly.
2. Backup rate (time) is too high, the patient is
breathing slower than the backup rate.
Remedy
1. Clean filters.
2. Unblock air inlet.
3. Lower room temperature.
4. Locate device away from heat source.
5. Lower heater setting.
1. Ensure proper contact is made (Fig. 3).
1. Adjust the inspiratory and expiratory trigger
sensitivity to adjust triggering for the patient.
2. Reduce Backup Rate setting.
O
2
O under normal use
2
O/l air
2
EN - 13