4.6 Variations in operating range
Quantity
Meas. range
Quantity
quantities
quantities
influenced
Battery voltage 7.5 to 10 V
Temperature
0...50°C
Ω
W var VA
Relative
10...90% HR
humidity
Ω
W var VA
Frequency
10 Hz...1 kHz
1 kHz...5 kHz
10 Hz...250 Hz
250 Hz...2.5 kHz
Position of
Position
conductor
on
in the
perimeter
jaws
internal
W var VA
(f ≤ 400 Hz)
jaws
Retentivity
0...600 peak
Adjacent
Conductor
conductor
in contact with
crossed by
external
a current
perimeter
W var VA
400 A
or
jaws
DC
RMS
Conductor
clamped by
0...400 A DC
the clamp
or RMS
Application
0...600 V DC
of a voltage
or RMS
on the clamp
Peak
1,4 to 3.5 limited A
AC
factor (1)
to 600 A peak
V
900 V peak
AC
0.7 and I ≥ 5 A
PF
(inductive and 0.5 and I ≥ 10 A
W var
0.2 and I ≥ 20 A
capacitive)
Rejection of
0...600 V/50 Hz
V
serial mode
0...400 A/50 Hz
A
in DC
Rejection of
V
AC
0...600 V DC
serial mode
W var VA PF
in AC
A
AC
0...400 A DC
W var VA PF
Rejection of
common
0...600 V/50 Hz
mode
W var VA PF
Influence of
external
0...400 A/m
magnetic
(50 Hz)
W var VA PF
field
Number of
manoeuvres
50000
opening of
W var VA PF
jaws
(1) The influence on quantities W, var, VA and PF is identical to that on the current
assuming sinusoidal voltage
Influence
Typical
Maxi
All
< 1 pt
0.2% L +1 pt
V
0.05% L/10°C
0.2% L /10°C +2 pt
A
0.3% L/10°C
0.5% L /10°C +2 pt
0.1% L/10°C
0.2% L /10°C +2 pt
0.25% L/10°C
0.5% L /10°C +2 pt
PF
< 1 pt
2 pt
Hz
0.03% L/10°C
0,1% L /10°C +2 pt
≤ 1 pt
V
0.1% L +1 pt
A
0.2% L
0.3% L +2 pt
0.2% L
0.3% L +2 pt
0.25% L
0,5% L +2 pt
PF
< 1 pt
1 pt
Hz
0.25% L
0.5% L +2 pt
1% L +1 pt
V
see curve
6% L +1 pt
1.5% L +1 pt
A
see curve
6% L +1 pt
A
0.7% L
1% L +1 pt
A
2 mA/A
3 mA/A
A
45 dB
40 dB
V
< 1 pt
1 pt
A
< 1 pt
1 pt
,
+
1% L
3% L +1 pt
AC
DC
,
+
1% L
3% L +1 pt
AC
DC
1% L +1 pt
0.5% L
3% L +1 pt
8% L +1 pt
50 dB
45 dB
DC
40 dB
35 dB
DC
,
+
AC
DC
> 60 dB
50 dB
,
+
AC
DC
> 50 dB
40 dB
V
< 1 pt
60 dB
A
0.07 A/100 V
0.1 A/100 V
< 1 pt
60 dB
A
70 dB
60 dB
A
0.3% L
1% +1 pt
43