Menu options
GB
Mode
On
Off
!
Store
Date
2
1
3
4
Time
Cal
Store
Date
diS
1
Time
Cal
Mode
Zero
Test
Mode
!
Store
Date
Prt
1
Time
Cal
50
Menu selections
Press the [MODE] key and hold.
Switch the unit on using the [ON/OFF] key.
Allow the 3 decimal points to be displayed before releasing the
[MODE] key.
The [!] key allows for selection of the following menu points:
diS recall stored data
1
Prt transmitting stored data
1
setting the date and time
2
3
user calibration
4
The selected menu is indicated by an arrow in the display.
diS – Recall of stored data
1
After confirming the selection with the [MODE] key the photometer
shows 125 data sets in the following format (automatically proceeds
every 3 seconds until result is displayed):
Number
n xx (xx: 16...1)
Year
YYYY (e.g. 2014)
Date
mm.dd (monthmonth:dayday)
Time
hh:mm (hourhour:minuteminute)
Test
Method
Result
x,xx
The [ZERO/TEST] key repeats the current data set.
The [MODE] key scrolls through all stored data sets.
Quit the menu by pressing [!] key.
Prt – Transmitting stored data - Bluetooth
1
The MD 110 has a Bluetooth
less transmission of data. Now it is possible to transmit stored results.
Bluetooth
4.0 is also known as Bluetooth
®
(Low Energy). Data is transmitted from the photometer as a .csv file.
Details on how information is transmitted from the photometer can
be found on www.lovibond.com. To receive the data, there are several
options on offer from the Tintometer
The App, AquaLX
, is available for mobile devices such as Smartphones
®
and Tablets and enables the user to manage and graphically chart the
received data. Both the data and charts can then be shared via email.
AquaLX
can be downloaded free of charge from the iTunes Store
®
for iOS
®
and from Google Play
4.0 interface which enables the wire-
®
Smart or Bluetooth
®
Group.
®
™
Store for Android
™
MD110_2 03/2018
®