4.5 POWER SUPLLY

Power supply :
Pile 9 V LF22/6LR61
Battery life :
> 100 heures à température ambiante
Automatic switching-off
15 minutes of non-use
time :
4.6 COMPLIANCE WITH INTERNATIONAL
STANDARDS
Application of safety rules as per standard
EN-61010-1-Ed.2:2001.1000V CAT-III -
Electrical safety :
600V CAT-IV. Pollution level 2. Double
insulation.
Electromagnetic
Compliant with standard EN-
compatibility :
61326/A2:2001Residential environment
Mechanical
Free fall: 1m (in accordance with
strength :
standardIEC-68-2-32)
Degree of protection
IP54 as per EN 60529
of the housing :

4.7 VARIATIONS IN DOMAIN OF USE

Quantity of influence
Range of influence
Temperature
0° C to + 50°C
Frequency
1 to 3 kHz
English
Influence
V AC : 0,5%L/10° C
mV DC : 0,5%L/10° C
(R>20 M ) :
0,5%L/10° C
µF (C>50µF)
5%L/10° C
V AC : 10%L + 1ct
44 - 117