2. Remove the dirty filter with a small, pointed object. Dis-
card the filter.
3. Replace with a clean filter. Additional filters are available
from your Graham-Field distributor.
4. Replace the filter cap.
Caution: Do not use cotton or other materials as a fil-
ter.
Caution: Do not wash or clean the filter.
Caution: Do not reuse the filter. Do not operate com-
pressor without a filter. Compressor damage can
result.
Caution: The disposable nebulizer kit is for single pa-
tient use only.
Storage
Keep the device and accessories dry. Avoid exposure to
direct sunlight.
TROUBLESHOOTING
If your Neb-u-Tyke Penguin Nebulizer Compressor fails to
function, consult the following Troubleshooting guide. If
the problem persists, consult your Graham-Field distribu-
tor.
Problem
Device doesn't
operate
Nebulization is
weak
Cause and Solution
Check if plug is properly fit into a grounded electrical outlet.
When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if filter is dirty; replace filter as needed.
John Bunn Neb-u-Tyke Penguin Nebulizer Compressor