5 Cable connections
5.5
Electronic corner correction
5.5.1
General information
Minebea Intec
Minebea Intec load cells are produced according to high quality standards and have
precisely adjusted output values.
Nonetheless, mechanical imbalances can cause impermissible corner load errors to arise,
which will need to be ofset by soldering in resistors.
The correct installation and accurate alignment of load cells are imperative for good
measurement results and signiicantly afect the behavior with corner loads. Therefore,
the installation and alignment of the load cell should always be checked irst if a corner
error is identiied.
If necessary, carry out mechanical height adjustment (see Installation manual of the load
cell).
Cable Junction Box PR 6130/04N
EN-15