3.0
Setting the main voltage and replacing
the fuses
Warning
• The fuse box must not be removed from the power
control switch until the mains cable has been
disconnected.
• The two main fuses must be identical and correspond to
the chosen voltage.
Locate the main fuses in the fuse box in the power control
switch and check the voltage of the power control switch
(on/off switch (230V/115V)). If the voltage of the power
control switch differs from the line voltage, you must adjust
the voltage of the power control switch.
Open the lid of the fuse box using a screwdriver.
Remove the fuse box.
Remove both fuses and insert two new fuses. These must be
identical and should correspond to the line voltage.
ITC-155:
ENGLISH
115V, 2AT = 105014 / 230V, 1AT = 105007