Switch On/Off
Off
On
Temperature Unit Setup
Temperature Setup
EC Reference Temperature Setup
EC Temperature Coefficient Setup
EC Standard Solution Setup
Note: Predefined standard solution: 10 μS/cm, 84 μS/cm, 146.5 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm. Please set the Standard Solution first before EC calibration.
EC Cell K Setup
Note: Please set the Cell K value first before EC calibration.
EC TDS Factor Setup
Calibration
Touch the Calibration icon on the
screen.
Touch Finish to complete
calibration.
Measurement
Touch the mode area on the
screen to select measurement
Insert the probe into a sample.
mode.
Switch EC/TDS/SAL/RES Mode
EC
Press the
button when you
finish to continue.
The calibration result will be
shown when you finish.
Measuring
TDS
SAL
Touch the value on the screen
to confirm your setting.
Calibrating
Completed
RES
Touch the value on the screen
to confirm your setting.
Touch the value on the screen
to confirm your setting.
Touch the value on the screen
to confirm your setting.
Print
Print:
Touch the Printer icon on
the screen to print.
Store/Recall Data
Recall measurement data:
Touch the Beaker icon on the
screen to recall.
Note: measurement data will
be automatically stored after
each operation.