General Setup (applies for all parameters)

(PaTH: H \ COnFIG \ Guided Setup)
h
10
Select the desired Channel for the guided setup and in the same line the parameter.
If auto is selected, M800 transmitter automatically recognizes the type of sensor.
The channel can also be fixed to a certain measurement parameter (parameter =
pH / ORP, UniCond2e, Cond4e, O2 hi, 02 lo, O2 Trace, O2 opt, O3 and Flow hi,
Flow lo, Flow Type 2), depending on the type of trans m itter.
Press the corresponding button to measurement M1 to configure the measurement.
NOTE: If the guided setup for a flow sensor has been selected, the calibration factor of
the sensor from the sensor label or certificate can be entered. Press therefore
the "CaL FaCTOR" button. For sensor types "High" and "Low" slope and offset
can be entered. In case of sensor "Type 2" the slope followed by a table of
"K" and "F" values can be entered.
assign the corresponding output signal Aout'X' to the measurement through press-
ing "YES".
Enter the Min Value, that corresponds with start point of the analog output range.
Enter the Max Value, that corresponds with end point of the analog output signal.
additional settings can be done by navigating to the next page of the menu.