Note:
-
Use the USB cable to connect the main instrument with the controller. The slot is located behind the
burette.
Turn return valve to position 1. In order to pour off the barrel turn return valve to direction 2
-
Easy70's system is ready to operate.
6. Operation Test
Ensure that the socket is properly grounded.
Plug the power cable into the system, then into the outlet. Make sure the power is on.
Refill the bottle with water and connect the burette to the top of the bottle, following the steps
previously explained.
Power on the instrument; the power button is in the back of the controller.
Wait until the self-check cycle is completed.
Do a complete cycle of filling in and dispensing before the first time work:
Suction: Hold the filling button
-
the burette is full, it will emit a double sound.
Dispensing: Hold dispensing button
-
liquid, it will emit a double sound.
The cycle of filling and dispensing needs to be done at least twice more.
If the previous action has been carried out correctly, the EASY 70 device is ready to operate.
- Please ensure that the adapter is fastened prior to each use.
- For perfect titration, please do not move or touch the instrument during
operation to avoid any damage to the EASY 70 device.
1
2
. The burette will automatically fill up with the liquid. Once
for fast titration. Once the burette has emptied all the
9