OxyGo Column Change Procedure (continued)
11. Insert column (metal tube) into the OxyGo
concentrator. Do not leave the column exposed;
it should be inserted into the OxyGo as
soon as the dust caps have been removed.
12. Push the column (metal tube) into the device
such that the column (metal tube) is fully seated
into the OxyGo concentrator. The spring
loaded latch button should fully return to the
closed position.
13. Repeat steps 10-12 to install the other column
(metal tube).
14. Connect the AC power supply cord to the OxyGo
and plug the power supply AC cord into an
electrical outlet. Do Not Power on the OxyGo
concentrator.
15. Press and hold the light button for 10 seconds and
screen will display "sieve reset" message. Release
button once message is displayed on screen.
16. Press the bell button once and screen will display "sieve reset successful".
17. Press the power
WARNING Do not use any columns other than those specified in this user manual.
The use of non-specified columns may create a safety hazard and/or
!
impair equipment performance and will void your warranty.
button to turn on the OxyGo, and use normally.
Cleaning, Care and Maintenance Chapter 6
Closed and locked
35