Lift off front cover and set aside.
Disconnect the sensor chip assembly from the block.
Gently lift the chip and the mounting post together out of the
block.
NOTICE! The sensor chip assembly is fragile. Handle
with care at all times. To avoid damaging the sensor with
a static electricity discharge, do not touch the sensor
circuitry, and do not place any tape or other material in
contact with the sensor circuitry.
Lift the heat exchange block straight upward off the base.
Note orientation of block for reassembly.
Inspect the heat exchange block carefully. If dust is
present, use compressed air or a HEPA-filtered vacuum
cleaner to gently clear the channels of the block.
07-01922C F532-120V 2015-03 Warranty 07-01939