2.4. MAINTENANCE OF THE APPLIANCE
Warning! : All maintenance operations shall only be performed by a technically
qualified service centre!
− To ensure correct and safe operation, the appliance must be inspected and serviced at
least once a year only. Maintenance includes also to control the components and tear
of pipes, feeding pipes, electrical components etc..
− It is advisable to replace worn components during maintenance operations to avoid the
need for other maintenance calls and unexpected failures.
− It is also advisable to apply for a maintenance contract with the customer.
2.4.1. POSSIBLE FAILURES AND THEIR ELIMINATION
Warning! : Only technically qualified service centres can perform the operations
described below!
Warning! : Before resetting the safety thermostat, it is always necessary to
eliminate the problem causing its activation!
Even a normal use of the appliance may cause operation inconveniences and
failures.
The most common problems are the following:
THE GREEN VOLTAGE CONTROL LIGHT DOES NOT LIGHT
Check that the power to the arrivals.
-
Check whether the main switch is activated.
-
Check the fuses F1 .
-
Lamp broken.
-
THE GREEN VOLTAGE CONTROL LIGHT
Check if switch with release coil IB is on
-
Lamp broken
-
Instruction manual for installation, maintenance and use
Centrifuge DRC 700V
(
AUX
.
:
)
:
DOES NOT LIGHT
DRC700V-GB-00