PROGRAMMING
Press Fn for 2 seconds in the start screen to program the device with 6 modes (PED1 - PED6).
Press OK to pass from a parameter to the next one.
It is possible to change the duration of all programs from 5' to 100' with steps of 5'.
Default duration 20'.
PED1:
It is possible to set pulse frequency and duration.
From 1 to 150 Hz with steps of 1 Hz.
From 50 to 250 µs with steps of 10 µs.
Default values: 100 Hz / 200 µs.
PED2:
It will be possible to set pulse duration with a fixed frequency of 100 Hz.
From 50 to 250 µs with steps of 10 µs.
Default values: 200 µs.
PED3:
It is possible to set frequency values.
From 1 to 150 Hz with steps of 1 Hz.
Default values: 100 Hz
Modulated pulse duration from 100 µs to 250 µs.
PED4:
It is possible to set pulse frequency and duration.
From 1 to 150 Hz with steps of 1 Hz.
From 50 to 320 µs with steps of 10 µs.
It is possible to change the following values: tup = 1 s; thold = 5 s; tdown = 1 s; toff = 5 s.
Default values: 30 Hz / 250 µs.
t
t
t
t
up
hold
down
off
PED5:
It is possible to set frequency values.
From 1 to 150 Hz with steps of 1 Hz.
Default values: 30 Hz
Modulated pulse duration from 100 µs to 320 µs.
Rev.4 22.06.2018
51