Technical Documentation; Electromagnetic Emissions - mebela Apgar Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 48

Technical Documentation

IEC 60601-1-2, Table 201

Electromagnetic Emissions

The Medela Apgar Timer is intended for use in the electromagnetic environment specified below.
The customer or the user of the Medela Apgar Timer should assure that it is used in such envi-
ronment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 60000-3-3
Warning – The Medela Apgar Timer should not be used adjacent to or stacked with other equipment. If ad-
jacent or stacked use is necessary, the Medela Apgar Timer should be observed to verify normal operation
in the configuration in which it will be used.
Compliance
Electromagnetic environment - guidance
Group 1
The Medela Apgar Timer uses RF energy only for their
internal function. Therefore, its RF emissions is very low
and is not likely to cause any interference in nearby
electronic equipment.
Class B
The Medela Apgar Timer is suitable for use in all es-
tablishments, including domestic establishments and
those directly connected to the public low-voltage
Class A
power supply network that supplies buildings used for
domestic purposes.
not applicable