Ortotecsa G.18907 Instrucciones De Uso Y Mantenimiento página 16

Tabla de contenido

Publicidad

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the infrared lamp is used exceeds the applicable RF compliance level above, the XXXX
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocation the infrared lamp.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE INFRARED LAMP
DISTANCIAS DE SEPARACIÓN RECOMENDADAS ENTRE LOS EQUIPOS PORTÁTILES Y MÓVILES
DE COMUNICACIONES DE RF Y EL BAÑO DE PARAFINA.
The paraffin bath is intended for use in a electromagnetic environment in which radiated RF disturbances are
controlled. The customer or user of the bath can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitter) and the paraffin bath as
recommended below, according to the maximum output power of the communication equipment.
El baño está previsto para usarse en un entorno electromagnético en el que se controlan las perturbaciones radiadas de
RF. El cliente o usuario puede ayudar a prevenir la interferencia electromagnética manteniendo una distancia mínima
entre el equipo portátil y móvil de comunicaciones de FR (transmisores) y el baño de parafina según se recomienda
debajo, conforme a la máxima potencia de salida del equipo de comunicaciones.
Rated maximum output
power of transmitter
W
Máxima potencia de
salida asignada del
transmisor
0,01
0,1
1
10
100
For transmitters rated at a maximum output power nor listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 to 80 MHz and 800 MHz, the separation distance is applied in the highest frequency range.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, object and people.
Ortotecsa,s.l. C/Peña de Francia,8
Tlf.:+34918719132
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,17 √P
0,12
0,38
1,2
3,8
12
28500 Arganda del Rey (Madrid), España.

Fax: +34918719323
email:
INSTRUCCIONES DE USO Y MANTENIMIENTO
USER AND MAINTENANCE MANUAL
Last Revision 12/03/2020
m
80 MHz to 800 MHz
d =1,17√P
0,12
0,38
1,2
3,8
12

info@ortotecsa.com
www.ortotecsa.com
800 MHz to 1 GHz
d=2,33√P
0,23
0,74
2,3
7,3
23