 Maintenance of the anode and the cathode
The sensitive part of the probe is composed by an anode and a cathode; the two elements are composed
by precious metals. Over time, these metals can be passivated by decreasing the efficiency of the probe, up
to the point that the probe does not calibrate.
In this case, remove the membrane and the passivation with very fine abrasive paper, gently scratching the
metal parts; wash everything with distilled water and proceed with the Replacement of Electrolyte.
Perform the Calibration of the sensor. If it does not work, proceed with the Replacement of the membrane.
 Replacement of the membrane
If the membrane has ripples or irregularities, it must be replaced with a new one.
Remove the membrane cap and replace it with a new and intact one.
With the new membrane, follow the procedure of the Replacement of Electrolyte.
If, even after these procedures, the probe does not calibrate, replace the probe.

Instrument Setup Menu

In measure mode, press key
Use the directional keys to move to SETTINGS P9.0 and access the menu by pressing the key
Move with the keys
The table below shows the setup menu structure for the general settings of the instrument; for each
program, there are the options that the user can choose and the default value:
Program
Description
P9.1
TEMPERATURE U.M.
P9.3
BACKLIGHT MODE
P9.4
BRIGHTNESS
P9.5
SLEEP MODE
P9.8
RESET
P9.9
AUTO POWER-OFF
P9.1 Unit of measurement for temperature
Access this setup menu to select the temperature unit to use:
°C -default-
°F
P9.3 Backlight Mode
Access this setup menu to select the contrast mode to use for the display backlight:
INDOOR (In) – Recommended if you use the device indoors.
OUTDOOR (Out) – Recommended if you use the device outdoors.
P9.4 Brightness
Access this setup menu to choose between three different levels of display brightness:
LOW – low
MEDIUM – medium
HIGH – high
Note: Keeping the display bright always adversely affects battery life
UM Serie Oxy7Vio EN rev.0 28.05.2020
to access the SETUP menu.
and
to select the program to access.
°C / °F
INDOOR – OUTDOOR
LOW – MEDIUM - HIGH
OFF – 2 MIN – 5 MIN
YES - NO
YES – NO
18
Options
Factory Default Settings
.
°C
INDOOR
MEDIUM
2 MIN
NO
YES