What to do if...

The display remains blank
when switched on
appears
Date/time is lost when
switched off
Password forgotten
Photometer does not
react

Error messages:

remove cell
lamp defective
no zero adjustment
cell error
cell invalid
method invalid
wrong method
E_0
E-1, E_2 or E_3
Release 06/2019
Spectroquant
Connect the photometer to the line power via the power pack.
In the case of battery operation: Battery empty, charging required
(approx. 5h); line operation is possible without restrictions during
charging time
Battery nearly empty. Charging required (see chapter
COMMISSIONING).
The backup battery of the real time clock is empty and has to be
replaced.
Send the photometer to the service department for this.
Inform the service department.
The connected printer is off line. Switch on the printer or pull out
the interface cable.
The message remove cell appears on the display although no cell is
inserted. Clean the cell shaft with a damp, lint-free cloth.
If the error message still appears, return the photometer to the
service department.
Replace the lamp (see chapter MAINTENANCE, CLEANING,
DISPOSAL).
No zero adjustment is stored in the meter for the cell. Perform zero
adjustment (see chapter ZERO ADJUSTMENT).
The rectangular cell is inserted incorrectly, or two cells are inserted.
Insert the cell correctly.
A cell type that is not allowed for the selected method was selec-
ted, e. g. a round cell for the reagent test.
No data is stored in the photometer for the selected method.
Update method data (see chapter UPDATING METHOD DATA, in the
FUNCTIONAL DESCRIPTION document [on CD-ROM]).
During a difference measurement, the method was changed bet-
ween the first and second measurement. During a difference
measurement, the method must remain identical.
Hardware error: Send the photometer to the service department.
Replace the lamp (see chapter MAINTENANCE, CLEANING,
DISPOSAL).
If the error message remains, send the meter to the service
department.
NOVA 60A - Operating Manual
®