Specifications; Maintenance; Changing Battery - Chauvin Arnoux C.A 1226 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

4. SPECIFICATIONS

Metrological specifications
Measurement
Resolution
range
Air velocity
0,15 to 3 m/s
0,01 m/s
3,1 to 30 m/s
Ambient
-20 to +80°C
temperature
Air flow
0 à 99 999 m3/h
Technical features
Conditions of use:
0 to 50°C; < 85 %RH
neutral gas
Storage conditions:
-20 to 80°C; 15 to 85 %RH
Power supply:
1 alkaline battery, 9V 6LR6
Device shuts off between 0 to 120 minutes
Protection IP 54
Dimensions and weight
Instrument:
147.7 x 70.6 x 34.7mm
Cable:
L = 2m
Probe:
Ø = 100mm, L = 296mm
Display:
4 lines
2 lines of 5 digits for value
2 lines of 5 digits for unit
50 x 34.9mm
Weight:
190g
Conformity
Electromagnetic Compatibility: complies with NF EN 61326-1.

5. MAINTENANCE

5.1
Cleaning
Clean the casing with a cloth and a little soapy water.
Clean off with a damp cloth.
5.2

Changing battery

-
Remove the battery cover
-
Insert a 9V battery
-
Replace the battery cover
Accuracy
Measurement
± 3% R + 0,1 m/s
m/s , fpm, km/h
0,1 m/s
± 1% R + 0,2 m/s
0,1°C
± 0,3% R + 0,25 °C
1 m3/h
± 3% R ± 0,03 * Surf.
m3/h, m3/s, L/s,
duct(cm²)
12
units
°C, °F
cfm

Publicidad

Tabla de contenido
loading

Tabla de contenido