The screen displays:
"Unit Overheat"
"Please Wait"
"Switch Off to"
"reset"
"menu"
The screen displays:
"Over Temp"
If any of these problems persist, the Grant Service Department can be contacted on
+44 (0) 1763 260811 or
XUB Series
Operating Manual
The Grant digital ultrasonic bath has detected that the
ultrasonic generator on the unit has operated above
the rated temperature and has disabled the ultrasonic
function. This also disables a cycle from starting. The
most likely cause is the fluid evaporating when the
level sensor is switched off. Switch the unit off, allow
the unit to cool for a few minutes, top up the liquid
above the level sensor and switch the unit on to reset
the alarm.
The Grant digital ultrasonic bath has detected that the
water in the bath is above 75ºC and has aborted the
cycle. Please allow the liquid to cool or drain some
and replace with cooler liquid.