6. Technical Specifications

Control box
Batteries
Charger
Motor
Hand Control
IP class
Material
Castors
Weight
Expected Service life
Max patient weight
Operating environment
Storage and transportation
environment
Height
Length
Width
Turning diameter
Materials in contact with
human skin
Applied parts:
P - patient,
C - caregiver
x
X
12
Linak, voltage 230V/24V, Max 1.5A, IPX6, Class II (Double insulated)
Linak, 24 V/1.2 Ah valve-regulated, sealed, lead accumulator (gel-type batteries,)
IPX5, Cable length 200mm
Max. 400 mA
Linak, DC 24 V, 4,3A, IP X6. Operation time 5% at maximum continous operation of 1
minute, Push: 2000N. Stroke 400mm +/-2 mm
Cable length coiled 600 mm, IPX5
IPX5
Steel
125/100/75mm
33 kg
8 years
240 kg / 529 Ibs
5°C to 40°C, relative humidity 20% to 80% non-condensing, and atmospheric
pressure 700 to 1060 hPa.
5 °C to 50 °C, relative humidity 20% to 80% non-condensing, and atmospheric.
Environment pressure 700 to 1060 hPa.
125 mm wheels model minimum 910 mm and maximum 1310 mm
100 mm wheels model minimum 890 mm and maximum 1290 mm
75mm wheels model minimum 860 mm and maximum 1260 mm
780mm
890mm
125 mm wheels: 1212mm, 100 mm wheels: 1183 mm, 75 mm wheels: 1183 mm
Materials that user and operator/caregiver are exposed to during usage of the pro-
duct for prolonged time:
Poroflex (synthetic leather cushions) P
Polyurethane (on PU cushions) P
Dryflex (handles) P
PA6 + 20% GF (brake lever) P
Hand control C
X
Armrest P
Handles P
Knee pad P
Foot rest support P
Foot rest support clamps C
Handles adjusting knobs C
Hand controller and its holder. C
Handles to adjust the armrest width. C
Hand control C
Charger C
Charger cable C