ImpORTANT pARTS OF yOUR DEvILBISS IgO BATTERy CHARgER
1.
Battery Slot
2.
Charge Connectors
3.
Control Panel
4.
External Power Light (Green)
5.
Service Required Light (Red)
6.
Charge Indicator Light (Green)
7.
External Power Input (on back)
8.
Service Connection (on bottom) Factory Use Only
9.
AC Power Cord (not shown)

OpERATINg yOUR IgO BATTERy CHARgER

1.
Place the iGo External Battery Charger on a flat, clean, dry location near a wall outlet.
2.
Ensure that the charger is properly connected to an AC power source with adequate voltage (100-240VAC). While the
charger is connected to an AC power source, the External Power Light will illuminate.
3.
Insert a DeVilbiss iGo Battery Pack (part number 306D-413) into the charger with the front of the battery pack facing
forward so that the power connectors on the back of the battery pack align with the charge connectors in the battery slot.
a.
While the battery is being charged, the green Charge Indicator Light will flash.
b.
When the battery is fully charged, the green Charge Indicator Light will stop flashing and change to a constant
green light.
4.
You may remove the battery pack at any point in the charge cycle.
5.
The battery can remain in the charger after it is fully charged as long as the charger is powered.
CAUTION–Do not leave the battery in the charger if the charger is not powered. The charger will discharge the battery if the charger is not externally powered.
NOTE–If a faulty battery is inserted into the charger, it may take up to 30 seconds for the charger to indicate the fault.

NORmAL OpERATION

Green Charge
Red Service
Light
Light
ON
ON
FLASHING
OFF
ON
OFF

CHARgE INDICATOR LIgHTS/TROUBLESHOOTINg

Green
Red
Charge
Service
Light
Light
ON
ON
OFF
OFF
OFF
ON
EN - 4
Indication
When power is first applied to the charger, both the green charge light
and the red service light should illuminate no longer than 3 seconds.
Battery is charging
Battery is fully charged
Green
External
Power
Light
Indication
ON
Green Charge Light and Red
Service Light remain lit after
the normal 3-second test at
start-up
ON
Charger does not detect an
installed battery
ON
Battery Error
Reasons
Charger Error
1. Battery not inserted
properly
2. Charger or battery error
Battery Error
2
1
3
8
Green External Power Light
ON- Receiving AC Power/ Normal
Operation
ON - AC Power/Normal Operation
ON - AC Power/Normal Operation
Remedy
Contact your DeVilbiss provider to repair / replace
unit
1. Reinsert battery per instruction guide
2. Contact your DeVilbiss provider to repair /
replace unit
Reinsert battery per instruction guide. If symptoms
persist contact your DeVilbiss provider to repair /
replace unit