Caring For Your Igo Battery Charger; Specifications - DeVilbiss iGo Manual De Instrucciones

Cargador de baterías externo para el sistema de oxígeno portátil
Ocultar thumbs Ver también para iGo:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 7
Green
Red
Charge
Service
Light
Light
OFF
FLASHES
FLASHES
FLASHES
FLASHES
ON
OFF
OFF

CARINg FOR yOUR IgO BATTERy CHARgER

WARNINg
Before attempting any cleaning procedures, unplug the unit from the wall outlet.
Battery slot and air vents should be kept free of dirt or contaminants.
Use a soft, damp cloth to clean the surfaces of the unit.

SpECIFICATIONS

Dimensions ........................................................................................................................................7.75" (19.69 cm) W x 4.75" (12.07 cm) D x 3.825" (9.72 cm) H
Input Power .........................................................................................................................................................................................................................100-240V~
Charge Time ........................................................................................................... Approximately 3 hours depending on ambient temperature and residual charge
IPX0 Rating
Operating Temperature............................................................................................................................................................................. 5˚C to 40˚C (41˚F to 104˚F)
Electrical Requirements ..................................................................................................................................................................................... 100-240V~, 50/60 Hz
Maximum input current rating (100-120V~) ...................................................................................................................................................................1.9A (rms)
Maximum input current rating (220-240V~) ...................................................................................................................................................................1.1A (rms)
Maximum charging output ............................................................................................................................................................................. 28V
Storage Conditions
Temperature: .................................................................................................................................................................................... -20°C to + 60°C (-4°F to 140°F)
Humidity .........................................................................................................................................................................................10% to 95% R.H., non-condensing
Barometric Pressure (tested at ~933 hPa) ......................................................................................................................................................... 500 hPa to 1060 hPa
Mode of operation .............................................................................................................................................................................................................. Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide
A-306CH
Green
External
Power
Light
Indication
ON
Battery failed voltage output
test
ON
Battery does not charge
ON
Battery does not charge
OFF
Charger does not detect AC
power source
Reasons
Battery Error
Charger's internal temperature
is too hot to continue the
charge cycle due to high
ambient temperature or fan
failure
Battery pack is too hot to
continue the charge cycle due
to high ambient temperature
or battery failure
1. Charger is not receiving
correct power
2. Power cord is not properly
connected to the charger
3. Power cord or charger is
faulty
Remedy
Reinsert battery per instruction guide. If symptoms
persist contact your DeVilbiss provider to repair /
replace unit
1. Ensure adequate airflow is available around the
charger
2. Ensure air vent holes are not occluded or
blocked
3. Place charger in a cooler location for 1 hour.
4. If symptoms persist after 1 hour in cooler
location, contact your DeVilbiss provider to repair /
replace unit
1. Remove battery pack from charger and place in
a cool location for 30 minutes.
2. Reinsert the cooled battery
3. If symptoms persist
a. insert a second battery, if one is available. If
second battery resolves issue, replace failed
battery
b. contact your DeVilbiss provider to repair /
replace unit
1. Ensure the charger power cord is fully inserted
into a live wall outlet
2. Check the power connection at the back of the
charger
3. If problem persists after verifying power source,
contact your DeVilbiss provider to repair / replace
unit
3.58A Max
EN - 5

Publicidad

Tabla de contenido
loading

Tabla de contenido