5. UV Light Operation
Overview
By using the 10-minute timer, sharable devices can be cleaned in a shorter amount of time than the traditional method of wiping the surfaces. Doing
so also prevents damage to devices caused by liquid chemicals.
Usage
Once the user opens and closes the doors on the unit, the internal
timer resets. The timer cycles for 10 minutes when the "START" button
is pressed.
1. Open the door of the storage cart.
2. Insert devices and ensure the unit is plugged in.
3. Close the door. The "DOOR CLOSED / READY" LED will illuminate.
Press the "START" button and the UV lights will turn on automatically.
4. If the door is opened before the timer finishes its 10-minute cycle,
the lights will turn off and the timer will be reset for the next
sanitizing cycle.
Note: Both storage cart doors must be closed before the start button is pressed.
The UV lights will not turn on if either door is open. This helps ensure
non-protected eyes or skin are never exposed directly to UV light.
Maintenance and Troubleshooting
• Check that the lamp connector is securely attached to lamp.
• Check that the fuse is in good working order. Replace if necessary.
• Check that proper power is being supplied to the device and that the switch lights when turned on.
• Lamps should be cleaned every six months to remove any dust accumulation on the bulbs that might reduce the effectiveness of the UVC light.
• Check if switch is working properly by opening and closing the cabinet doors and see the light turning off/on. The timer should operate for
10 minutes before shutting off.
Warnings and Safety
• A broken UV lamp is extremely unlikely to have any negative impact on your health. If a lamp breaks, ventilate the room for 30 minutes and remove
the parts while wearing gloves. Put the parts in a sealed plastic bag and take it to your local waste facility for recycling. Do not use a vacuum
cleaner.
• CAUTION! The lamps are a Risk Group 3 Ultra Violet product. These lamps emit high-power UV radiation that can cause severe injury to the skin
and eyes. Avoid skin and eye exposure to unshielded product. Use only in an enclosed environment that shields users from radiation.
Dimensional Drawing
6. Specifications
Model
Cabinet Dimensions (H x W x D)
Unit Weight
Load Capacity
Power Requirement
Operating Temperature
Operating Humidity
Product
TUV 8W
FAM/10X25BOX
CSCSTORAGE1UVC
40.78 x 24.30 x 30.33 in. / 1036 x 617 x 770 mm
198.42 lb. / 90.0 kg
UV LIGHT
CYCLE "ON"
(10 min)
A (max.)
B (max.)
288.3 mm
295.4 mm
36.68 x 20.20 x 30.33 in. / 932 x 513 x 770 mm
250 lb. / 113 kg Stationary or Rolling
Input: 120V AC, 60 Hz, 12 Amps
32° to 104°F / 0° to 40°C
5 to 95% RH, Non-Condensing
7
DOOR CLOSED
DOOR CLOSED
/ "READY"
/ "READY"
START BUTTON
B (min.)
C (max.)
293 mm
302.5 mm
CSCSTORAGE2UVC
138.89 lb. / 63.0 kg
D (max.)
16 mm