All manuals and user guides at all-guides.com
IMPORTANT: Start by reading the Easypump C-bloc RA Pump Instructions For Use, then continue with this document.
ILLUSTRATION AND NOMENCLATURE
Figure 1
1. E-Clip
2. Fill Port
3. Easypump C-bloc RA Pump
4. Clamp
5. Filter
6. Flow Controller (tape to skin)
PRIMING THE ADMINISTRATION SET
• Open clamp.
• Remove tubing cap to prime pump (up to 15
minutes). When tubing is fully primed, close
clamp and replace tubing cap until ready for
use.
STARTING THE INFUSION
• Connect catheter to pump tubing. Open
clamp to begin infusion.
• Tape Flow Controller to skin to ensure flow
rate accuracy. Avoid contact with cold
therapy. Do not tape over filter.
CAUTION
• Temperature: The Easypump C-bloc RA Flow
Controller should be in direct contact with
the skin (88˚F/31˚C). Temperature will affect
solution viscosity, resulting in faster or slower
flow rate. If Easypump C-bloc RA is used
with the Flow Controller at room temperature
(68˚F/20˚C), flow rate may decrease by
approximately 25%.
See Table 1 on back for Flow Rate Accuracy
and Delivery Time information.
Rx Only =
CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
U.S. and Foreign Patents Pending.
® C-bloc and Easypump are registered trademarks of I-Flow Corporation.
1
➏
4
➊
➋
➌
➍
➎