CAUTION: Do not fill less than minimum or exceed maximum fill volume (see below).
TABLE 1 – Delivery Time Information
Easypump RA
Nominal Flow Rate (ml/hr)
Nominal Fill Volume (ml)
Maximum Fill Volume (ml)
Retained Volume (ml)
hours from refrigerator
hours from freezer
Approximate Delivery Time
24 h
30 h
1.3 d
36 h
1.5 d
48 h
60 h
2.5 d
72 h
96 h
120 h
135 h
5.5 d
When filled to nominal volume, flow rate accuracy is ±15% of the labeled infusion rate
when infusion is started 0-8 hrs. after fill and delivering nominal saline as the diluent at
88˚F (31˚C) against a back pressure of 40 cm of water.
CAUTION: Filling the pump less than nominal, increases flow rate. Filling the pump
greater than nominal, decreases flow rate.
270-5
5
270
335
≤ 8
12
24
1 d
150
2 d
255
290
3 d
330
4 d
5 d
6 d
270-10
300-6
10
6
270
300
335
335
≤ 8
≤ 8
Time to Reach Room Temperature
12
12
24
24
Fill Volume (ml)
250
170
290
325
235
285
335
5
400-5
400-8
5
8
400
400
550
550
≤ 15
≤ 15
12
12
24
24
230
225
300
270
400
310
350
440
400-10
10
400
550
≤ 15
12
24
280
370
460